Merilnik motnosti meglice za pivo - HI847492
The HI847492 Haze Meter is specially designed for measurement of haze in beer quality analysis. The instrument is based on a state-of-the-art optical system which guarantees accurate results, assures long term stability, and minimizes stray light and color interferences. It also compensates for variations in intensity of the lamp, although proper calibration ensures measurements are fully validated, comparable, and in compliance with regulatory requirements. The 25 mm round cuvettes composed of special optical glass guarantee the repeatability of turbidity measurements.
Features at-a-glance
ASBC Compliant Measurement - The HI847492 meets and exceeds the requirements of the ASBC method for haze measurements. Measurements are taken from beer samples that have been chilled and degassed, and results are reported in Formazin Turbidity Units (FTU).
Calibration –A two, three, or four-point turbidity calibration can be performed by using the supplied (
AMCO AEPA-1 Primary Turbidity Standard - The AMCO AEPA-1 supplied standards are recognized as a primary standard by the USEPA. These non-toxic standards are made of styrene divinylbenzene polymer spheres that are uniform in size and density. The standards are reusable and stable with a long shelf life.
Fast Tracker™ - For users taking measurements throughout their brewery, the HI847492 is equipped with Fast Tracker™ - Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows brewers to record the time and location of a specific measurement or series of measurements using iButton® tags on specific tanks for quick and easy readings. Each iButton® tag contains a computer chip with a unique identication code encased in stainless steel.
GLP Data – The HI847492 features complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. Data includes calibration points, date, and time.
Data Logging – Up to 200 measurements can be stored in the internal memory and recalled at any time.
Data Transfer – For further storage or analysis options, logged data can be downloaded to a Windows compatible PC using the USB port and the HI92000 software.
Backlit Display – A backlit LCD display provides an easy to understand, user-friendly interface. Displayed codes guide the user step-by-step through routine operation and calibration.
Significance of Use
Beer clarity is a parameter constantly controlled in the brewery. Clarity can be affected by haze, or insoluble or semi-soluble particulate matter which is small enough to form a colloidal suspension in beer, typically less than 2 mm. These particles scatter transmitted light and are observed as a degradation in the brilliance of the beer. To assure consistent product quality, the brewmaster needs more than visual inspection.
Several substances can cause haze in beer, but the most frequently encountered problem is due to a cross-linking of polyphenols and proteins. These materials exist in equilibrium in beer and manifest themselves as a haze when they combine to form insoluble colloids. A range of stabilization treatments is available for avoiding haze problems, and the product has to be controlled in several steps during the brewing process, particularly after filtration and before the beer enters the bright tanks.
The HI847492 Haze Meter reports the measurements in FTU (Formazin Turbidity Units). FTU units are equal to NTU units (Nephelometeric Turbidity Units). The conversion table between these measurement units is shown below:
Conversion Table
|
FTU/NTU/FNU |
EBC |
ASBC |
HELM |
1 FTU 1 NTU 1 FNU |
1 |
0.25 |
17.25 |
10 |
1 EBC |
4 |
1 |
69 |
40 |
1 ASBC |
0.058 |
0.014 |
1 |
0.579 |
1 HELM |
0.1 |
0.025 |
1.725 |
1 |
Grade |
EBC |
ASBC |
Brilliant |
0.0 to 05 |
0.0 to 34.5 |
Almost Brilliant |
0.5 to 1.0 |
34.5 to 69 |
Very Slightly Hazy |
1.0 to 2.0 |
69 to 138 |
Slightly Hazy |
2.0 to 4.0 |
138 to 276 |
Hazy |
4.0 to 8.0 |
276 to 552 |
Very Hazy |
> 8.0 |
> 552 |
The HI847492-11 AMCO AEPA-1 standards ensure that measurements are traceable to a primary reference materials. These standards are used for calibration and performance verification of the turbidity meter.
Specifications
sku |
HI847492-01 |
FTU Range |
0.00 to 9.99 FTU |
FTU Resolution |
0.01 FTU from 0.00 to 9.99 FTU |
FTU Accuracy |
±2% of reading plus 0.05 FTU |
Range Selection |
Automatic |
Turbidity Repeatability |
±1% of reading or 0.02 FTU, whichever is greater |
Stray Light |
< 0.1 FTU |
Turbidity Light Detector |
Silicon Photocell |
Turbidity Method |
Ratio Nephelometric Method. |
Calibration |
Two, three or four point calibration |
Light Source |
LED @ 580nm |
Display |
60 x 90mm backlit LCD |
LOG Memory |
200 records |
Serial Interface |
USB 1.1 |
Environment |
0 to 50°C (32 to 122°F); max 95% RH non-condensing |
Power Supply |
4 x 1.5V AA alkaline batteries or AC adapter |
Auto Shut-Off |
After 15 minutes of non-use |
Dimensions |
224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) |
Weight |
512 g (18 oz.) |
Warranty |
2 years |
Supplied with Certificate of Analysis:
- Lot number
- Expiration date
- Standard value @ 25 °C
- Reference meter NIST traceable
Provided storage containers:
- Light tight
- Protects from accidental breakage